Fan bank of a calibration wind tunnel
Calibration: air velocity

Air velocity calibration from 0.1 to 70 m/s

UKAS-accredited calibration for vane, thermal and ultrasonic anemometers and Pitot tubes, carried out in wind tunnels in Lymington by Labcal, the Gill Group's calibration laboratory.

What is air velocity calibration?

Air velocity calibration compares an anemometer or Pitot tube with a reference velocity in a controlled wind tunnel. The result is the instrument's error at each test speed and the uncertainty of that measurement, so readings used for safety checks, commissioning or research can be shown to be traceable.

Labcal uses laser Doppler anemometry and reference meter methods, so the reference velocity is measured without disturbing the airflow. Low-speed work for cleanroom, fume cupboard and HVAC instruments runs in one tunnel; a second accredited facility handles larger instruments and higher speeds, up to 50 m/s.

Handheld vane anemometer showing an air velocity reading
Vane anemometer

Accredited from cleanroom airflow to 50 m/s

Every range below sits within Labcal's UKAS accreditation. The second facility also covers ultrasonic anemometers.

Air velocity calibration accredited ranges and best measurement capability
Instrument or quantityRangeExpanded uncertainty (k=2)
Thermal anemometers0.1 to 5 m/sQ[0.19 %, 0.030 m/s]
Thermal anemometers5 to 38 m/sQ[0.18 %, 0.030 m/s]
Vane anemometers, up to 125 mm diameter0.3 to 5 m/sQ[0.24 %, 0.030 m/s]
Vane anemometers, up to 125 mm diameter5 to 38 m/sQ[0.22 %, 0.040 m/s]
Pitot tubes2 to 5 m/sQ[0.19 %, 0.030 m/s]
Pitot tubes5 to 38 m/sQ[0.18 %, 0.030 m/s]
Large and ultrasonic anemometers, Pitot tubes5 to 50 m/sQ[1.7 %, 0.16 m/s] to Q[1.7 %, 0.88 m/s] by speed band

Ranges and best measurement capability from the UKAS schedule of accreditation for laboratory 0625, issue 055, 18 August 2026. Q[a, b] means the two terms are combined in quadrature. The uncertainty on your certificate depends on the instrument under test.

Check your instrument with us

The instruments behind airflow decisions

01

Vane anemometers

Handheld and fixed vane instruments used for duct, grille and general airflow checks, up to 125 mm diameter in the low-speed tunnel.

02

Thermal and hot wire anemometers

Low-velocity probes used for cleanroom, laminar flow and fume cupboard face velocity work, accredited from 0.1 m/s.

03

Pitot tubes

Pitot-static tubes used with a pressure readout for duct traverses and higher-velocity measurements.

04

Ultrasonic anemometers

Two and three-axis ultrasonic wind sensors from a wide range of manufacturers, calibrated in the larger facility up to 50 m/s.

05

Instruments with displays

Instruments are calibrated with their own handheld or integrated display, so the certificate reflects the reading the user actually sees.

06

Legacy instruments

Older instruments still in service can be calibrated. If an instrument is not listed here, ask and the laboratory will confirm suitability.

Typical makes include Airflow, ATP, Fluke, Kestrel, Kimo, RS, Testo and TSI.

From your bench to a certificate and back

01

Tell us what needs calibrating

Use the form on this page with the instrument, the quantity and the points you need. If you are moving from another laboratory, your current certificate is enough for a like-for-like quote, usually within 24 hours.

02

Send it to Lymington

Instruments come to the Labcal laboratory at Ampress Park, Lymington. Calibration is laboratory based rather than on site. Courier collection can be arranged for regular and high-volume customers.

03

Calibration against traceable references

Turnaround averages 8 working days from the day the instrument arrives, with a 3-day fast track available by arrangement. Where an instrument can be adjusted, it is recorded as found, adjusted and recorded again as left.

04

Certificate and return

A UKAS certificate states the measured values, the expanded uncertainty at k=2 and the traceability to national standards, with as-found and as-left results where adjustment was made.

Tell us what needs calibrating

The instrument, how many and the points you need are enough to start. If you are switching laboratory, mention it and we will ask for your current certificate so the quote matches it like for like.

This field is for validation purposes and should be left unchanged.
We send the quote here.
Sending more than one? List them all here.
What do you need?(Required)
Switching from another laboratory? We will ask for your current certificate so the quote matches it like for like.
Please enter a number greater than or equal to 1.
Calibration is priced per unit.
Month and year if you know it.

UKAS accredited. Labcal, laboratory 0625, ISO/IEC 17025:2017.

We aim to respond in one working day. Turnaround averages 8 working days from arrival, with a 3-day fast track by arrangement.

Switching laboratory? Send your current certificate for a like-for-like quote.

Frequently asked questions

Labcal's UKAS accreditation for air velocity runs from 0.1 m/s to 50 m/s. Thermal anemometers are accredited from 0.1 m/s, vane anemometers from 0.3 m/s and Pitot tubes from 2 m/s, all up to 38 m/s in the low-speed tunnel. A second facility covers 5 m/s to 50 m/s, including large and ultrasonic anemometers.

Yes. Ultrasonic anemometers are named on Labcal's UKAS schedule for the facility that covers 5 m/s to 50 m/s. That includes instruments from a wide range of manufacturers. The achievable uncertainty depends on the instrument, its size and how it is mounted in the test section.

Laser Doppler anemometry measures air velocity optically by tracking particles in the flow, so nothing is placed in the airstream to disturb it. Labcal uses it as one of its reference methods, which suits low-uncertainty work such as calibrating reference instruments or research anemometers.

The calibration interval for an anemometer should be set by your quality system, based on how critical the measurement is, the accuracy you need and how stable the instrument has been at previous calibrations. Annual calibration is common for cleanroom and HVAC commissioning instruments, and some users calibrate before critical projects.

Yes. A UKAS air velocity certificate records the measured error at each calibrated velocity together with the expanded measurement uncertainty at k=2. Where an instrument is adjusted, the certificate shows the as-found and as-left results so you can see what changed.

Talk to an engineer

Tell us what you need to measure, where and under what conditions. We’ll help you work out the appropriate approach.