Calibration: temperature

Temperature calibration from -30 °C to +140 °C

UKAS-accredited calibration for resistance thermometers, thermocouples, digital thermometers, probes and data loggers, carried out in Lymington by Labcal, the Gill Group's calibration laboratory.

What is temperature calibration?

Temperature calibration places a sensor in a stable, known temperature and compares its reading with a traceable reference thermometer. The certificate states the error and the measurement uncertainty at each calibrated point, so a probe, logger or indicator can be relied on for quality records, storage monitoring or process control.

Labcal calibrates sensors in dry blocks and fluid baths using reference meter methods, and calibrates temperature probes in air in an environmental chamber. Temperature indicators and simulators can also be calibrated electrically, by simulating thermocouple and resistance thermometer signals across much wider ranges.

Temperature probes with a handheld digital thermometer
Temperature probes

Sensors, indicators and loggers on one schedule

Sensor calibration runs from -30 °C to +140 °C in baths and blocks. Electrical calibration of indicators extends far beyond that, which suits instruments that read thermocouples at furnace or process temperatures.

Temperature calibration accredited ranges and best measurement capability
Instrument or quantityRangeExpanded uncertainty (k=2)
Resistance thermometers, in block or bath-30 to +140 °C0.10 °C
Base metal thermocouples, in block or bath-30 to +140 °C0.25 °C
Indicators with resistance thermometer probes-30 to +70 °C0.10 °C
Indicators with resistance thermometer probes70 to 140 °C0.15 °C
Indicators with base metal thermocouple probes-30 to +140 °C0.25 °C
Temperature probes in air, environmental chamber-5 to 10 °C0.22 °C
Temperature probes in air, environmental chamber10 to 60 °C0.15 °C
Electrical: base metal thermocouple types T, N, K and J-200 °C to the type limit, up to +1,372 °C0.50 °C
Electrical: noble metal thermocouple types S and R0 to 1,760 °C0.90 °C
Electrical: PRT simulation, Pt50 to Pt1000-200 to 850 °C0.16 to 0.36 °C by range

Ranges and best measurement capability from the UKAS schedule of accreditation for laboratory 0625, issue 055, 18 August 2026. Q[a, b] means the two terms are combined in quadrature. The uncertainty on your certificate depends on the instrument under test.

Check your instrument with us

Every point in the chain from sensor to display

01

Resistance thermometers

PRTs and RTDs, calibrated on their own or with the indicator they are read through.

02

Thermocouples

Base metal thermocouples in blocks and baths, and thermocouple indicators electrically across their full type range.

03

Digital thermometers

Handheld and bench thermometers, with fixed or interchangeable probes.

04

Temperature data loggers

Loggers used for cold chain, storage, transport and environmental monitoring.

05

Transmitters, indicators and controllers

Instruments that convert or display a temperature, including electrical calibration with simulated inputs.

06

Probes in air

Air temperature probes calibrated in an environmental chamber from -5 °C to 60 °C.

Typical makes include Digitron, Fluke, Omega, RS, RS PRO, Testo, TSI and WIKA.

From your bench to a certificate and back

01

Tell us what needs calibrating

Use the form on this page with the instrument, the quantity and the points you need. If you are moving from another laboratory, your current certificate is enough for a like-for-like quote, usually within 24 hours.

02

Send it to Lymington

Instruments come to the Labcal laboratory at Ampress Park, Lymington. Calibration is laboratory based rather than on site. Courier collection can be arranged for regular and high-volume customers.

03

Calibration against traceable references

Turnaround averages 8 working days from the day the instrument arrives, with a 3-day fast track available by arrangement. Where an instrument can be adjusted, it is recorded as found, adjusted and recorded again as left.

04

Certificate and return

A UKAS certificate states the measured values, the expanded uncertainty at k=2 and the traceability to national standards, with as-found and as-left results where adjustment was made.

Tell us what needs calibrating

The instrument, how many and the points you need are enough to start. If you are switching laboratory, mention it and we will ask for your current certificate so the quote matches it like for like.

This field is for validation purposes and should be left unchanged.
We send the quote here.
Sending more than one? List them all here.
What do you need?(Required)
Switching from another laboratory? We will ask for your current certificate so the quote matches it like for like.
Please enter a number greater than or equal to 1.
Calibration is priced per unit.
Month and year if you know it.

UKAS accredited. Labcal, laboratory 0625, ISO/IEC 17025:2017.

We aim to respond in one working day. Turnaround averages 8 working days from arrival, with a 3-day fast track by arrangement.

Switching laboratory? Send your current certificate for a like-for-like quote.

Frequently asked questions

Labcal's UKAS accreditation covers temperature sensors from -30 °C to +140 °C in blocks and baths, and temperature probes in air from -5 °C to 60 °C. Electrical calibration of temperature indicators and simulators extends from -200 °C up to 1,760 °C, depending on the thermocouple type or resistance thermometer being simulated.

Electrical calibration of a temperature indicator checks the instrument by feeding it a simulated thermocouple or resistance thermometer signal rather than placing a probe in heat. It verifies the indicator, including reference junction compensation for thermocouples, across temperatures that would be impractical to generate in a bath.

Calibrating a probe and indicator together gives the error of the system as it is used, which is usually what matters for a reading on a process or in a store. Calibrating them separately is useful where probes are swapped between indicators. Labcal's schedule covers both arrangements.

The calibration interval for a temperature instrument depends on how critical the measurement is, regulatory requirements and how the instrument has performed at previous calibrations. Annual calibration is common for critical instruments, but shorter or longer intervals may suit the application and should be set by your quality system.

ISO 9001:2015 clause 7.1.5 requires measuring equipment to be suitable for its purpose and, where measurement traceability is required, calibrated against standards traceable to national or international standards. UKAS-accredited calibration is a common way to demonstrate this, although other justified routes can be acceptable.

Talk to an engineer

Tell us what you need to measure, where and under what conditions. We’ll help you work out the appropriate approach.