Tunnel ventilation and air flow

In a tunnel, air flow is a safety system.
Measure it like one.

Ultrasonic wind sensors for tunnel ventilation control, at the portals, in the ducting and along the bore, with a tunnel mode built for measuring flow along a single axis.

0
Moving parts to seize, ice or wear
3
Positions measured: portal, duct and bore
4-20mA
Single channel tunnel mode output
1985
Building ultrasonic wind sensors in the UK

The ventilation system acts on the number, so the number has to be right

A tunnel ventilation system does three jobs. It keeps air quality inside the bore acceptable during normal running, it holds temperature within limits, and in a fire it controls where the smoke goes. All three are driven by a measurement of air flow, and the control system acts on that measurement without questioning it.

That is what makes a failed sensor dangerous rather than inconvenient. A mechanical anemometer that has seized does not report a fault. It reports still air, and still air is a plausible reading inside a tunnel. The fans respond to a number that is no longer true.

The environment does the seizing. Tunnel air carries exhaust particulates, brake and tyre dust, salt dragged in from the road, and a constant wash of moisture, and the bore is one of the least convenient places on an asset to reach for maintenance. A sensor in a duct or on a tunnel ceiling is a lane closure, a permit and a night shift.

Infrastructure sensing map

Every infrastructure decision starts at a measurement point

A transport corridor measures across the whole asset. Wind at the bridge deck and the tower, at the crane jib and the exposed viaduct. Road-level conditions, lineside exposure, portal pressure and the water level in the drainage below.

Explore 16 measurement points to see what is being measured, where the measurement happens and which sensing technology fits the application.

Get expert advice

Replacing a propeller anemometer on an existing control system?Tunnel mode, single channel 4-20mA output, mounting positions and traceable calibration. Direct from our engineering team.

Request information

Flow along the bore, not wind across a field

Measuring inside a tunnel is a different problem from measuring in the open. The air is constrained to one axis, it is pushed by traffic as much as by the fans, and what the control system needs is a signed value: how fast, and which way along the bore.

That is why the sensors used here run in a tunnel mode. Instead of reporting a compass direction, the sensor reports flow along one axis with a sign, so a single value tells the control system both the speed and whether the air is moving towards the north portal or the south. On a single channel 4-20mA output, 12mA is set as zero flow, with the range below it representing flow one way and the range above it flow the other.

The practical consequence is that an ultrasonic sensor can take over from an existing propeller anemometer without the control system being rewritten. Many tunnels were commissioned with propeller units feeding exactly that signal, and the replacement can present the same signal on the same wiring.

Measurement also happens in more than one place. Portals carry the exposed cross-wind problem for high-sided vehicles approaching at speed. Ducting carries the extraction flow. The bore itself carries the flow the smoke control strategy depends on.

Three places, three different questions

At the portal, the question is cross-wind on vehicles entering and leaving, which is a two-axis measurement in open air and the same problem a bridge or an exposed approach carries.

In the ducting and along the bore, the question is flow rate and direction on one axis, mounted to a side wall or a ceiling where a conventional head would foul on the structure and where nobody wants to return.

Portal, duct and bore are separate specifications

View wind sensors

The measurement that matters most is the one taken during an incident

In a fire, the strategy is to control where the smoke travels. In larger tunnels that means driving the flow towards zero in the zone around the incident so that smoke does not spread along the bore ahead of evacuating people.

Reaching and holding near-zero flow is a closed loop, and it depends on a sensor that is still reporting accurately at low flow rates and in hot, smoke-laden air. This is the case that should set the specification, not normal running.

Specify for the incident, not the average day

Discuss your requirements

A seized anemometer does not report a fault. It reports calm.

Every argument for solid-state measurement in a tunnel comes back to that sentence. A cup or propeller anemometer degrades quietly: bearings stiffen, the starting threshold creeps up, and the readings drift low long before anything stops. Inside a bore, nobody is watching it to notice.

An ultrasonic sensor has nothing to seize. It times sound pulses between fixed transducers, so there is no starting threshold, no bearing wear and no routine part replacement. In an asset where access is a night closure, that is a maintenance argument as much as a measurement one.

What a tunnel installation uses

Two ultrasonic ranges cover the positions a tunnel needs, with accredited calibration behind the measurement.

Wind and air flow measurement

Gill WindSonic 2-axis ultrasonic anemometers
Volume installation

WindSonic

2-axis ultrasonic, no moving parts

The sensor most tunnel installations are built on. Low cost enough to place at several points along an asset, with a tunnel mode for single-axis flow and a single channel 4-20mA option for replacing an existing propeller unit.

Tunnel mode : Reports signed flow along one axis rather than compass direction
Retrofit : Single channel 4-20mA output, 12mA at zero flow
Models : WindSonic 60, WindSonic 75, WindSonic M and an SDI-12 variant

View WindSonic

Gill WindObserver precision ultrasonic anemometers
Harsh duty

WindObserver

Precision 2-axis ultrasonic, stainless steel head

Where the duty is harder than a standard installation: a stainless steel open head that clears itself of debris, an optional de-icing system for exposed portals, and a hazardous-area variant where the tunnel carries a classified zone.

Construction : Stainless steel open head, built for exposed and contaminated air
Portals : Optional de-icing for approaches that ice in winter
Models : WindObserver 65, 70, 75 and 90, plus an intrinsically safe variant

View WindObserver

Calibration and verification

Traceable air velocity calibration, for equipment from any manufacturerIndependent, UKAS-accredited calibration covering air velocity and the other core parameters used in tunnel instrumentation, whoever made the sensor.

Explore calibration

Not sure which product fits this application?Tell us where it will be installed, what it has to survive and the system the data has to reach. Our team will narrow the range down with you.

Get expert advice

Where the measurement is taken

Road tunnels

Ventilation and pollutant control

Holding air quality inside the bore during normal running, and driving fans to clear carbon monoxide so that maintenance crews can work in the tunnel safely.

Rail and metro

Piston effect and extraction

Trains push air ahead of them through a confined bore. Measurement separates traffic-driven flow from the flow the ventilation system is producing.

Portals

Cross-wind on high-sided vehicles

The exposed approach either side of a tunnel is where a high-sided vehicle meets a sudden change in wind loading. That is a two-axis open-air measurement.

The output has to suit the system already in the tunnel

Tunnel control systems are long-lived and were commissioned around whatever the sensors of the day provided. A replacement that forces a change to the SCADA or the fan control logic is a much larger project than a sensor swap.

The ultrasonic ranges provide analogue and digital outputs, including the single channel 4-20mA tunnel mode that mirrors a propeller anemometer signal, so an upgrade can stay a sensor upgrade.

Match the signal, keep the control system

Read the application note

What the specification looks like in practice

Wind sensors have been supplied into tunnel applications worldwide for many years, in the portals, in the air ducting and on tunnel side walls and ceilings, which is where the tunnel mode came from in the first place.

Solid-state measurement removes the failure mode that matters in a confined asset: there is no bearing to stiffen and no starting threshold to creep, so a sensor that is working reports flow and a sensor that is not reports a fault rather than calm.

The sensors are built in the UK, and the measurement behind them can be verified against an independent, UKAS-accredited calibration for air velocity, which is the evidence a tunnel operator needs when the ventilation strategy has to be defended.

Ready to specify?Talk through approvals, mounting, power and integration with our team before the specification is fixed.

Discuss your requirements

Questions about tunnel ventilation measurement

Measuring inside a tunnel

Tunnel mode makes the sensor report air flow along a single axis with a sign, rather than reporting a compass direction. Inside a bore the air is constrained to one axis anyway, and what the control system needs to know is how fast it is moving and which way along the tunnel.

Gill engineering insightIt is also what makes a ceiling or side wall installation practical, because the sensor no longer has to be oriented to a compass reference.

Yes, where the existing unit feeds a single channel 4-20mA signal. That signal sets a centre point, usually 12mA, as zero flow, with the range below representing flow one way along the bore and the range above flow the other way. An ultrasonic sensor can be configured to present the same signal, so the control system sees what it has always seen.

Gill engineering insightThis is the difference between a sensor replacement and a control system project, and it is worth confirming the existing signal convention before specifying anything.

Three positions are typical and they answer different questions. Portals measure cross-wind on vehicles entering and leaving, in open air and on two axes. Ducting measures the extraction flow. The bore itself measures the flow the ventilation and smoke control strategy acts on, usually from a side wall or ceiling mount.

Safety and operation

Because the strategy is to control where the smoke goes. In larger tunnels the extraction fans are driven to bring flow towards zero in the zone around the incident, so that smoke does not travel along the bore ahead of people evacuating. That is a closed loop around a flow measurement, and it needs a sensor that is still accurate at low flow rates.

Gill engineering insightSpecify for this case rather than for normal running. The incident is the condition the measurement exists for.

In road tunnels the flow is deliberately altered so that maintenance crews can work in the bore without being exposed to carbon monoxide from vehicle exhaust. The ventilation system is clearing the working area, and the measurement confirms that it has.

Because the risk sits at the portal rather than inside. A high-sided vehicle leaving a tunnel moves from still, sheltered air into an exposed approach in a very short distance, and the change in wind loading is abrupt. Portal measurement supports the warning systems and speed restrictions that manage it.

Specifying and maintaining

Tunnel air carries exhaust particulates, brake and tyre dust, road salt and persistent moisture. On a mechanical anemometer that combination attacks the bearings, which raises the starting threshold and biases readings low. An ultrasonic sensor has no bearings and no routine part replacement, which is why it is the usual choice where access is difficult.

Gill engineering insightAccess cost is the hidden half of the specification. A sensor in a bore is a lane closure, a permit and a night shift every time it is touched.

WindSonic covers most installations, including multiple measuring points along one asset, and carries the tunnel mode and the 4-20mA retrofit option. WindObserver suits harder duty: a stainless steel open head for heavily contaminated air, optional de-icing for exposed portals, and an intrinsically safe variant where a classified zone is involved.

Where the ventilation strategy has to be evidenced, yes. Air velocity is one of the core parameters covered by independent, UKAS-accredited calibration, and that accreditation applies to instrumentation from any manufacturer, not only to the sensors described here.